Quantification of Triazine Pesticides by QuEchERS and LC/MS/MS
LC Conditions
Column
Synergi 2.5 µm Fusion-RP 100 Å, LC Column 50 x 2 mm, Ea
Brand
Synergi
Part No
00B-4423-B0
Phase Name
Fusion-RP
Elution Type
Gradient
Mobile Phase
A: Water with 5 mM Ammonium Formate
B: Metanol with 5 mM Ammonium Formate
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
8
10
90
3
14
10
90
4
15
80
20
5
20
80
20
Flow Rate
250.00 mL/min
Temperature
25°C
Detection
Tandem Mass Spec (MS-MS) @ (500°C)
Detection Info
Order Items Used in This Application
- 1. Desethylatrazine
- 2. Desisopropyl atrazine
- 3. Cyanazine
- 4. Hexazinone
- 5. Simazine
- 6. Metribuzin
- 7. Atrazine-d5
- 8. Ametryn
- 9. Metazachlor
- 10. Prometon
- 11. Sebutylazine
- 12. Atrazine
- 13. Propazine
- 14. Terbuthylazine
- 15. Prometryn
- 16. Terbutryn
- 17. Metolachlor
Similar Applications
No Similar applications found.