For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Fast Pain Panel using Kinetex 2.6 u Biphenyl 50x3.0mm
22121
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Antiemetics
Clinical
Pharmaceutical/Biopharmaceutical
Forensics/Toxicology
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Fast Pain Panel using Kinetex 2.6 u Biphenyl 50x3.0mm
Analytes
1Morphine
2Oxymorphone
3Hydromorphone
4Amphetamine
5Naloxone
6Methamphetamine
7Codeine
8MDA
9Oxycodone
10Naltrexone
11Hydrocodone
12MDMA
13MDEA
14Norfentanyl
15Tramadol
16Benzoylecgonine
17Meperidine
18Meprobamate
19Norbuprenorphine
20Fentanyl
21Buprenorphine
22Flurazepam
23Carisoprodol
24PCP
25Propoxyphene
26Sufentanil
276-MAM
28Midazolam
29Normeperidine
30EDDP
31Methadone
32Lorazepam
33Clonazepam
34Norpropoxyphene
35Oxazepam
36Hydroxyalprazolam
37Nordiazepam
38Flunitrazepam
39Temazepam
40Alprazolam
41Diazepam
Compound Name
Morphine
CID
5288826
Molecular Formula
C17H19NO3
Molecular Weight
285
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 22121)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% formic acid
B: Methanol with 0.1% formic acid
Gradient Profile:
Step No.Time(min)%A%B
109010
22.50100
33.50100
43.519010
559010
Flow Rate:
0.7 mL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) (40°C)

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase