EPA 8081 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)

EPA 8081 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)
GC Conditions
Official Method

EPA 8081

Column

Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea

Part No

7HM-G028-51

Phase Name

ZB-CLPesticides-1

Injection

Splitless @ 250°C

Oven Program

120 °C to 200 °C @ 45 °C/min to 230 °C @ 15 °C/min to 330 °C @ 30 °C/min for 2 min

Carrier Gas

Constant Flow Helium, 3.9 mL/min

Detection

Electron Capture (ECD) @ (330°C)

Detection Info
Sample Notes

Analytes are 250 ppb in hexane

Order Items Used in This Application

Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
7HM-G028-51

Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea

View Product

  • 1. 2,4,5,6-TCMX (Surr)
  • 2. a-BHC
  • 3. g-BHC
  • 4. b-BHC
  • 5. d-BHC
  • 6. Heptachlor
  • 7. Aldrin
  • 8. Heptachlor epoxide
  • 9. trans-Chlordane
  • 10. cis-Chlordane
  • 11. Endosulfan I
  • 12. 4,4--DDE
  • 13. Dieldrin
  • 14. Endrin
  • 15. 4,4'-DDD
  • 16. Endosulfan II
  • 17. 4,4'-DDT
  • 18. Endrin aldehyde
  • 19. Endosulfan sulfate
  • 20. Methoxychlor
  • 21. Endrin ketone
  • 22. Decachlorobiphenyl (Surr)