EPA 8081 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)
GC Conditions
Official Method
EPA 8081
Column
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Part No
7HM-G028-51
Phase Name
ZB-CLPesticides-1
Injection
Splitless @ 250°C
Oven Program
120 °C to 200 °C @ 45 °C/min to 230 °C @ 15 °C/min to 330 °C @ 30 °C/min for 2 min
Carrier Gas
Constant Flow Helium, 3.9 mL/min
Detection
Electron Capture (ECD) @ (330°C)
Detection Info
Sample Notes
Analytes are 250 ppb in hexane
Order Items Used in This Application
- 1. 2,4,5,6-TCMX (Surr)
- 2. a-BHC
- 3. g-BHC
- 4. b-BHC
- 5. d-BHC
- 6. Heptachlor
- 7. Aldrin
- 8. Heptachlor epoxide
- 9. trans-Chlordane
- 10. cis-Chlordane
- 11. Endosulfan I
- 12. 4,4--DDE
- 13. Dieldrin
- 14. Endrin
- 15. 4,4'-DDD
- 16. Endosulfan II
- 17. 4,4'-DDT
- 18. Endrin aldehyde
- 19. Endosulfan sulfate
- 20. Methoxychlor
- 21. Endrin ketone
- 22. Decachlorobiphenyl (Surr)
Similar Applications
No Similar applications found.