Oxytocin Extracted Ion Chromatograms, QCL (0.500 ng/mL)
LC Conditions
Column
Luna 3 µm PFP(2) 100 Å, LC Column 50 x 2 mm, Ea
Brand
Luna
Part No
00B-4447-B0
Phase Name
PFP(2)
Elution Type
Gradient
Mobile Phase
A: 0.1%FA in water
B: 0.1%FA in 50:50 MeOH:Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
0.2
90
10
3
0.5
70
30
4
3.5
70
30
5
3.51
30
70
6
4
30
70
7
4.01
90
10
8
5
90
10
Flow Rate
0.45 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (40°C)
Detection Info
Sample Preparation Details
Description
Strata-X-CW 33 µm Polymeric Weak Cation, 30 mg / well, 96-Well Plates, 2/Pk
Part No
8E-S035-TGB
Condition 1
1 mL Methanol at 1 mL/min
Condition 2
1 mL Water at 1 mL/min
Sample Pretreatment
Strata - X - CW is selected, which works for both compounds
Load
Load 300.00 µL Sample diluted in 4% H3PO4 in water
Wash 1
1 mL 4% H3PO4 in water at 1 mL/min
Wash 2
1 mL 30% MeOH in water at 1 mL/min
Dry
3-5 min
Elution 1
0.75 mL 1% TFA in 50:50 MeOH: Acetonitrile at 0.75 mL/min
Elution 2
0.75 mL 1% TFA in 50:50 MeOH: Acetonitrile at 0.75 mL/min
Evaporation
Evaporate to dryness
Reconstitute
100 µL 1% TFA in 80:20 MPA:MPB
Order Items Used in This Application
8E-S035-TGB
Strata-X-CW 33 µm Polymeric Weak Cation, 30 mg / well, 96-Well Plates, 2/Pk
- 1. Oxytocin
- 2. ARG-Vasopressin