Analysis of Parabens from Baby Wash on Kinetex 5u EVO C18 100x3.0
LC Conditions
Column
Kinetex 5 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, Ea
Brand
Kinetex
Part No
00D-4633-Y0
Phase Name
EVO C18
Elution Type
Gradient
Mobile Phase
A: 10% acetonitrile in water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
4
70
30
3
5
60
40
4
6
60
40
Flow Rate
1.50 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
Analytes 1. Methyl paraben 2. Ethyl paraben 3. Propyl paraben 4. Butyl paraben Novum SLE 12cc Tube 8B-S138-KDG Novum SLE 12cc Tube Extraction Procedure Sample pre-treatment 0.5 ± 0.05 g spiked Aveeno Baby (Wash & Shampoo) sample Add 1.25 mL acetone Add 0.6 ml 1% Nacl solution Vortex for 30 secs Wait 30 secs till the emulsion dissipates Centrifuge for 10 mins at 3000 rpm Collect the supernatant and proceed to load Sample loading "Load the sample from pre-treatment step above onto the Novum SLE cartridge and apply a short and gentle pulse of vacuum (~ 5" of Hg for 5-10 sec or until the sample has completely entered the media). "Wait for 5 minutes. Elution "Dispense 10 mL of ethylacetate onto the SLE media and allow the solvent to elute by gravity for 10 mins and collect the eluant. "Apply vacuum at 5"-10"of Hg to complete the extraction. NOTE: Prolonged application of vacuum will result in elution of body wash out of the SLE media and into the final extracted solvent. Dry down "Evaporate the final extract to complete dryness under slow stream of N2 at 45°C Reconstitute "Reconstitute the dry residue in 0.5 mL of methanol "Centrifuge for 5 mins at 6000 rpm "Collect supernatant to shoot on HPLC
Order Items Used in This Application
- 1. Methyl paraben
- 2. Ethyl paraben
- 3. Propyl paraben
- 4. Butyl paraben