Carcinogenic amines on Kinetex 2.6 µm EVO C18

Carcinogenic amines on Kinetex 2.6 µm EVO C18
LC Conditions
Column

Kinetex 2.6 µm EVO C18 100 Å, LC Column 150 x 2.1 mm, EA

Brand

Kinetex

Part No

00F-4725-AN

Phase Name

EVO C18

Elution Type

Gradient

Mobile Phase

A: 10 mM ammonium acetate/ ACN (95:5)

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

10

5

95

Flow Rate

0.40 mL/min

Temperature

50°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

Forced degradation protocol: Reconstitute 1 mg of cefaclor neat standard with 1.0 mL of 1 M HCl. Heat the solution at 60 ºC for 6 hours then neutralize with 1.0 mL of 1 M NaOH. Refrigerate until ready to inject.

Order Items Used in This Application

Kinetex 2.6 µm EVO C18 100 Å, LC Column 150 x 2.1 mm, EA
00F-4725-AN

Kinetex 2.6 µm EVO C18 100 Å, LC Column 150 x 2.1 mm, EA

View Product

  • 1. 4-Methyl-m-phenylenediamine
  • 2. o-Anisidine
  • 3. o-Toluidine
  • 4. Benzidine
  • 5. 4-Chloroaniline
  • 6. 3,3'-Dimethoxybenzidine
  • 7. 3,3'-Dimethylbenzidine
  • 8. 2-Naphthylamine
  • 9. 2,4,5-Trimethylaniline
  • 10. 4,4'-methylenedio-toluidine
  • 11. 4-Aminobiphenyl
  • 12. 3,3'-Dichlorobenzidine
  • 13. 2,2'-Dichloro-4,4'-methylenedianiline
  • 14. o-Aminoazotoluene
Similar Applications