ETG/ETS in urine using Luna Omega Polar C18 100x2.1mm

ETG/ETS in urine using Luna Omega Polar C18 100x2.1mm
LC Conditions
Column

Luna Omega 1.6µm Polar C18 100Å, LC Column 100 x 2.1 mm, EA

Brand

Luna Omega

Part No

00D-4748-AN

Phase Name

Polar C18

Elution Type

Gradient

Mobile Phase

A: 10mM ammonium formate with 0.1% formic acid

B: Acetonitrile with 0.1% formic acid

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

0

2

1

50

50

3

1.1

100

0

4

5

100

0

Flow Rate

300.00 mL/min

Temperature

25°C

Detection

Tandem Mass Spec (MS-MS) @ (450°C)

Detection Info

Order Items Used in This Application

Luna Omega 1.6µm Polar C18 100Å, LC Column 100 x 2.1 mm, EA
00D-4748-AN

Luna Omega 1.6µm Polar C18 100Å, LC Column 100 x 2.1 mm, EA

View Product

  • 1. Urine interference
  • 2. ETS
  • 3. ETG
Similar Applications