EtG / EtS from Urine on Luna Omega 5u Polar C18, 50x4.6mm
LC Conditions
Column
Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 4.6 mm, EA
Brand
Luna Omega
Part No
00B-4754-E0
Phase Name
Polar C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in H2O
B: 0.1% Formic Acid in MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
98
2
2
2
40
60
3
2.01
10
90
4
3
10
90
5
3.01
98
2
6
4
98
2
Flow Rate
1.00 mL/min
Temperature
25°C
Detection
Tandem Mass Spec (MS-MS) @ (550°C)
Detection Info
Sample Notes
MMA (117.0-->72.9; 117.0-->54.9) MMA-D3 (120.1-->75.9)
Sample Preparation Details
Description
Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S038-TAK
Condition 1
1 mL Methanol
Condition 2
1 mL 0.1% Acetic Acid in Water
Load
Load 100.00 µL Sample diluted in 0.1% Acetic Acid in Water
Wash 1
0.5 mL 50:50 Methanol:Water
Dry
5-10 min under high vacuum
Elution 1
0.6 mL 2% Ammonium Hydroxide in Methanol
Elution 2
0.6 mL 2% Ammonium Hydroxide in Methanol
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile Phase A
Order Items Used in This Application
- 1. Interference
- 2. Ethyl Sulfate (EtS)
- 3. Ethyl Glucuronide (EtG)
Similar Applications
No Similar applications found.