EtG / EtS from Urine on Luna Omega 5u Polar C18, 50x4.6mm

EtG / EtS from Urine on Luna Omega 5u Polar C18, 50x4.6mm
LC Conditions
Column

Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 4.6 mm, EA

Brand

Luna Omega

Part No

00B-4754-E0

Phase Name

Polar C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid in H2O

B: 0.1% Formic Acid in MeOH

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

98

2

2

2

40

60

3

2.01

10

90

4

3

10

90

5

3.01

98

2

6

4

98

2

Flow Rate

1.00 mL/min

Temperature

25°C

Detection

Tandem Mass Spec (MS-MS) @ (550°C)

Detection Info
Sample Notes

MMA (117.0-->72.9; 117.0-->54.9) MMA-D3 (120.1-->75.9)

Sample Preparation Details
Description

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk

Part No

8B-S038-TAK

Condition 1

1 mL Methanol

Condition 2

1 mL 0.1% Acetic Acid in Water

Load

Load 100.00 µL Sample diluted in 0.1% Acetic Acid in Water

Wash 1

0.5 mL 50:50 Methanol:Water

Dry

5-10 min under high vacuum

Elution 1

0.6 mL 2% Ammonium Hydroxide in Methanol

Elution 2

0.6 mL 2% Ammonium Hydroxide in Methanol

Evaporation

Evaporate to dryness

Reconstitute

200 µL Mobile Phase A

Order Items Used in This Application

Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 4.6 mm, EA
00B-4754-E0

Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 4.6 mm, EA

View Product

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk
8B-S038-TAK

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk

View Product

  • 1. Interference
  • 2. Ethyl Sulfate (EtS)
  • 3. Ethyl Glucuronide (EtG)
Similar Applications

No Similar applications found.