FAMEs In Palm Oil by GC/FID on Zebron ZB-FAME
GC Conditions
Column
Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea
Part No
7KG-G033-10
Phase Name
ZB-FAME
Injection
Split 100:1 @ 240°C
Oven Program
180 °C isothermal
Carrier Gas
Constant Flow Helium, 1.2 mL/min
Detection
Flame Ionization (FID) @ (240°C)
Detection Info
Sample Notes
Recommended Accessories: Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Weigh 0.2 g of sample into a glass test tube 2. Add 3.8 mL of hexane and vortex 3. Condition SPE tube (Strata® Si-1, 2 g/12 mL, Part No.: 8B-S012-KDG) under vacuum with 6 mL of hexane; do not let bed go dry 4. Load sample with light vacuum 5. Elute with 5 mL of hexane:ethyl acetate (87:13) 6. Evaporate under nitrogen at 35 °C 7. Reconstitute with 4 mL of hexane 8. Add 200 µL of 2 M potassium hydroxide/methanol solution and vortex 9. Leave at room temperature for 5 minutes 10. Add 2 mL of deionized water and vortex 11. Leave to separate until upper layer becomes clear 12. Extract upper layer for GC analysis
Sample Preparation Details
Pretreatment Note
To reconstituted sample, add 200 µL of 2 M potassium hydroxide in methanol, cap tube and vortex. Wat 5 minutes. Add 2 mL Milli-Q® water, vortex. Allow solution to settle then transfer top layer to Q-sert vial for GC analysis.
Condition 1
6 mL Hexane
Load
Load 3.80 mL Sample diluted in Pretreated Sample
Elution 1
5 mL Hexane/Ethyl Acetate (87:13)
Evaporation
Evaporate to dryness
Reconstitute
400 µL Hexane
Order Items Used in This Application
- 1. C12:0
- 2. C14:0
- 3. C16:0
- 4. C16:1 cis 9
- 5. C18:0
- 6. C18:1 cis 9
- 7. C18:2 cis 9,12
- 8. C18:3 cis 9,12,15
- 9. C20:0
- 10. C20:1 cis 11
Similar Applications