FAMEs In Canola Oil by GC/MS on Zebron ZB-FAME

FAMEs In Canola Oil by GC/MS on Zebron ZB-FAME
GC Conditions
Column

Zebron ZB-FAME, GC Cap. Column 30m x 0.25mm x 0.2 µm, Ea

Part No

7HG-G033-10

Phase Name

No Phase

Injection

Split 50:1 @ 240°C

Oven Program

100 °C for 2 min to 214 °C @ 3 °C/min

Carrier Gas

Constant Flow Helium, 1.0 mL/min

Detection

Mass Spectrometer (MS) @ (250°C)

Detection Info
Sample Notes

Recommended Accessories: Recommended Liner: Zebron PLUS Single Taper Z-Liner, 4 mm ID Liner Part No.: AG2-0A13-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Weigh out 25 mg of oil into a 5 mL screw cap glass vial 2. Add 2 mL of BF3-methanol to sample vial 3. Heat vial at 60 °C for 15 minutes with periodic vortexing 4. Remove from heat and allow to attain room temperature 5. Add 1 mL water followed by 1 mL hexane 6. Vortex for 1 minute 7. Remove upper organic layer 8. Dry hexane layer over anhydrous sodium sulfate 9. Transfer to autosampler vial and inject on GC

Order Items Used in This Application

Zebron ZB-FAME, GC Cap. Column 30m x 0.25mm x 0.2 µm, Ea
7HG-G033-10

Zebron ZB-FAME, GC Cap. Column 30m x 0.25mm x 0.2 µm, Ea

View Product

  • 1. C16:0
  • 2. C18:1 cis 9
  • 3. C18:2 cis 9,12
  • 4. C18:3 cis 9,12,15