TCAs on Kinetex 2.6u Polar C18, 50x2.1mm (100 ng/mL)
LC Conditions
Column
Kinetex 2.6 µm Polar C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4759-AN
Phase Name
Polar C18
Elution Type
Gradient
Mobile Phase
A: 10 mM ammonium formate in water
B: 0.1% formic acid in methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
55
45
2
2
35
65
3
2.5
5
95
4
3
5
95
5
3.1
55
45
6
4.5
55
45
Flow Rate
0.50 mL/min
Temperature
40°C
Detection
Tandem Mass Spec (MS-MS) @ (600°C)
Detection Info
Order Items Used in This Application
- 1. Nortriptyline
- 2. Protriptyline
- 3. Desmethyldoxepin
- 4. Desipramine
- 5. Amitriptyline
- 6. Doxepin
- 7. Imipramine
- 8. N-Desmethylclomipramine
- 9. Clomipramine