For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Extraction of Uracil and Dihydrouracil from Human Serum Using Strata-X PRO SPE Using Kinetex 2.6 μm PS C18 (ESI +Ve mode)
25777
Separation Mode: Reversed Phase
Reversed Phase
PS C18
HPLC
Clinical
Clinical Research
Therapeutic / Clinical

Extraction of Uracil and Dihydrouracil from Human Serum Using Strata-X PRO SPE Using Kinetex 2.6 μm PS C18 (ESI +Ve mode)

Analytes
15,6 Dihydrouracil (UH2)
2Uracil
3Uracil 1,3-15N2 (+Ve IS)
Details
LC Conditions (App ID: 25777)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
PS C18
System:
Agilent Technologies 1260 Infinity Agilent Technologies 1260 Infinity
Sample Note:
Add 100 µL Serum (doubly stripped) to tube. Add 900 µL of chilled (~00C) ACN to the serum while vortexing. Centrifuge samples at 6000 rpm for 5 minutes and transfer supernatant Analyte diluted with initial mobile phase