Gradient Optimization for Intact Mass LC Methods- Gradient Program Length
LC Conditions
Column
bioZen 2.6 µm WidePore C4, LC Column 100 x 2.1 mm, Ea
Brand
Biozen LC
Part No
00D-4786-AN
Phase Name
WIDEPORE C4
Elution Type
Gradient
Mobile Phase
A: 0.1 % Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0.9
95
5
2
7.5
10
90
3
7.8
10
90
4
8
95
5
5
8.2
10
90
6
8.3
10
90
7
8.4
95
5
8
12
95
5
Flow Rate
0.50 mL/min
Temperature
80°C
System
Waters ACQUITY UPLC I-Class Waters ACQUITY UPLC I-Class
Detection
LC/UV (DAD, PDA)
Detection Info
Sample Notes
Remicade
Order Items Used in This Application
Similar Applications