GC-MS/MS Separation of Heptachlorodibenzofurans (HpCDF) on ZB-Dioxin GC column with 5m Zebron Guardian
GC Conditions
Column
Zebron ZB-Dioxin w/GUARDIAN, 5 m, GC Cap. Column 60 m x 0.25 mm x 0.20 µm, Ea
Part No
7KG-G045-10-GGA
Phase Name
ZB-Dioxin
Injection
Splitless @ 330°C
Inlet Liner
AG2-0E02-05
Oven Program
to 220°C @ 30°C/min 220°C for 15.5 min to 240°C @ 2°C/min 240°C for 4.5 min to 270°C @ 5°C/min 270°C for 5 min to 330°C @ 15°C/min 330°C for 7 min
Carrier Gas
Constant Flow Helium, 1.1 mL/min
System
Agilent Technologies 7890 Unknown
Detection
GC/MS/MS @ (340°C)
Detection Info
Order Items Used in This Application
AG2-0E02-05
Zebron Plus Liner for Agilent & Thermo, 2mm ID Straight Wool in Middle, 5/Pk
7KG-G045-10-GGA
Zebron ZB-Dioxin w/GUARDIAN, 5 m, GC Cap. Column 60 m x 0.25 mm x 0.20 µm, Ea
- 1. 1,2,3,4,6,7,8-HpCDF
- 2. 1,2,3,4,7,8,9-HpCDF