GC-HRMS Separation of Hexa Furan using Zebron ZB-Dioxin GC column
GC Conditions
Column
Zebron ZB-Dioxin, GC Cap. Column 60m, 0.25mm, 0.20um, Ea
Part No
7KG-G045-10
Phase Name
ZB-Dioxin
Injection
Splitless @ 280°C
Inlet Liner
AG2-0E00-05
Oven Program
190°C for 3.5 min to 230°C @ 8°C/min 230°C for 15 min to 290°C @ 6.7°C/min 290°C for 5 min to 315°C @ 21°C/min 315°C for 9 min
Carrier Gas
Constant Flow Helium, 1.3 mL/min
System
HP 7890 HP GC/MS System (Unknown)
Detection
GC/HRMS @ (300°C)
Detection Info
Order Items Used in This Application
- 1. 1,2,3,4,7,8-HxCDF
- 2. 1,2,3,6,7,8-HxCDF
- 3. 1,2,3,7,8,9-HxCDF
Similar Applications