Levothyroxine and Impurities F and G according to Ph.Eur. Monograph 0401 on Luna Omega 3 µm C18 150 x 4.6 mm @1.32 mL/min
LC Conditions
Official Method
Ph.Eur. M.401
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: Dissolve 1.97 g of phosphoric acid in water and dilute to 2 L with water
B: Dissolve 1.97 g of phosphoric acid in acetonitrile and dilute to 2 L with acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
70
30
2
10
70
30
3
40
20
80
4
50
20
80
Flow Rate
1.32 mL/min
Temperature
25°C
System
Shimadzu Shimadzu Nexera UHPLC
Detection
LC/UV (DAD, PDA) @ 225 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Levothyroxine
- 2. Impurity F
- 3. Impurity G
Similar Applications