LC-MS/MS Pesticides ( simazine ) - No Filter in Cumin Exacts for the comparison of with/without Filter Vial
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 3.0 mm, Ea
Brand
Luna Omega
Part No
00F-4784-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid and 10mM Ammonium Acetate in water
B: 0.5% Formic Acid in Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
0.1
95
5
3
0.5
40
60
4
3.5
25
75
5
8
15
85
6
10.5
5
95
7
12
5
95
8
12.1
95
5
9
18
95
5
Flow Rate
0.40 mL/min
Temperature
20°C
System
Agilent Technologies 1100 Series SCIEX 5500 QTRAP
Detection
LC/MS/MS
Detection Info
Sample Notes
a) Prepared 10mg/mL simazine solutions with other 17 pesticide standards b) Take sample to prepare a 10x Diluted mixture by placing 100uL of the existing mix into a 900uL of a 1:2 Water:Acetonitrile mix. c) Place 30uL of the 10x diluted mixture into 270 uL of the cumin powder supernatant for final analysis.
Order Items Used in This Application
- 1. simazine
- 2. simazine
Similar Applications