USP Ezetimibe Organic Impurities Standard Solution 0.25mg/mL on Kinetex 5u F5 150x4.6mm on Agilent 1260 Quaternary UHPLC system
LC Conditions
Column
Kinetex 5 µm F5 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4724-E0
Phase Name
F5
Elution Type
Gradient
Mobile Phase
A: Water
B: Acetonitrile
C: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
66
24
2
37
66
24
3
60
40
50
4
70
40
50
5
80
10
80
6
90
10
80
7
90.1
66
24
8
100
66
24
Flow Rate
2.00 mL/min
Temperature
30°C
System
Agilent Technologies 1260 Infinity Agilent Technologies 1260 Infinity
Detection
LC/UV (DAD, PDA) @ 248 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. Desfluoroaniline analog
- 2. Ezetimibe�diastereomers
- 3. o-Fluorobenzene isomer
- 4. Ezetimibe
Similar Applications