USP Atazanavir Sulfate (Assay, Organic Impurities –Procedures 1 and 2, Limit of Atazanavir Related Compound A) Atazanavir system suitability solution-Zoomed scale. USP- Organic Impurity-Procedure-1
LC Conditions
Column
Luna Omega 3 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00F-4784-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 2.73 g/L of monobasic potassium phosphate in water. Adjusted pH with diluted Orthophosphoric acid to pH 3.50. Solution A(MP-A): Acetonitrile and Buffer (25:75) v/v.
B: 2.73 g/L of monobasic potassium phosphate in water. Adjusted pH with diluted Orthophosphoric acid to pH 3.50. Solution B(MP-B) : Acetonitrile and Buffer (75:25) v/v.
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
2
5
100
0
3
45
0
100
4
50
100
0
5
60
100
0
Flow Rate
1.00 mL/min
Temperature
25°C
System
Waters ACQUITY UPLC H-Class Waters ACQUITY UPLC H-Class
Detection
UV-Vis Abs.-Diode Array (PDA) @ 215 nm (nanometers)
Detection Info
Order Items Used in This Application
- 1. (S,R,S,S)-Diastereomer
- 2. (R,S,S,S)-Diastereomer
- 3. Atazanavir
Similar Applications