Sulfa Drug Screen - USP Method
LC Conditions
Official Method
USP L10
Column
Luna 3 µm CN 100 Å, LC Column 100 x 4.6 mm, Ea
Brand
Luna
Part No
00D-4254-E0
Phase Name
CN
Elution Type
Isocratic
Mobile Phase
A: Hexane with 0.1% Triethylamine
B: Methylene chloride:Methanol (80:20) with 0.1% Triethylamine
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
60
40
Flow Rate
4.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, CN 4 x 3.0mm, 10/Pk Part No.: AJ0-4305 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Sulfapyrazine
- 2. Sulfamerazine
- 3. Sulfadiazine
- 4. Sulfanilamide
- 5. Sulfathiazole
- 6. Sulfanilic acid
- 7. Sulfasalazine
- 8. Sulfaguanidine
Similar Applications