Amino Acids - OPA Derivatives
LC Conditions
Column
Luna 3 µm C18(2) 100 Å, LC Column 100 x 4.6 mm, Ea
Brand
Luna
Part No
00D-4251-E0
Phase Name
C18(2)
Elution Type
Isocratic
Mobile Phase
A: 0.1M Na2HPO4, anhydrous and 0.1 mM Na2-EDTA dihydrate in water, pH 6.38
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
70
30
Flow Rate
1.20 mL/min
Temperature
40°C
Detection
Fluorescence (FLUOR) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Aspartate
- 2. Glutamate
- 3. Asparagine
- 4. Histidine
- 5. Serine
- 6. Glutamine
- 7. Arginine
- 8. Glycine
- 9. Threonine
- 10. Taurine
- 11. Alanine
- 12. gamma-Aminobutyric acid
Similar Applications