Particle Size 8 |
Recommended Use Sugar alcohol analysis according to the USP procedure, monosaccharides and sugar alcohol analysis, cellobiose, glucose, xylose, arabinose, and mannose and other cellulose products |
| Phase | Description | Applications | Bio-Rad® Aminex® | Supelco® Supelcogel™ | Waters® Sugar-Pak™ | Transgenomic® CARBOSep™ | Sepax® Carbomi® |
|---|---|---|---|---|---|---|---|
| RPM-Monosaccharide (USP L34 packing) | 8 % cross-linked LEAD | – 100 x 7.8 mm dimensions available for sugar alcohol analysis according to the USP procedure – Monosaccharides and sugar alcohol analysis – Cellobiose, glucose, xylose, arabinose, and mannose and other cellulose products |
HPX-87P 125-0098 |
Supelcogel Pb |
N/A | CARBOSep COREGEL-87P |
Carbomix Pb |
| Counter Ion Analyte | RPM Pb2+ |
|---|---|
| Adonitol (Ribitol) | 20.15 |
| D-Altrose | 15.82 |
| D-(-)-Arabinose | 16.47 |
| D-(+)-Cellobiose | 11.00 |
| D-(+)-Digitoxose | 15.32 |
| Dulcitol | 33.25 |
| Meso-Erythritol | 19.82 |
| D-(-)-Fructose | 17.71 |
| L-(-)-Fucose | 16.19 |
| D-(+)-Galactose | 14.94 |
| Gentiobiose | 10.53 |
| D-(+)-Glucose | 14.94 |
| Inositol | 18.87 |
| Isomaltose | 11.28 |
| Lactose | 11.89 |
| Lactulose | 13.95 |
| D-Lyxose | 16.66 |
| Counter Ion Analyte | RPM Pb2+ |
|---|---|
| D-Maltose | 11.59 |
| Maltotriose | 11.02 |
| Maltulose | 12.40 |
| D-Mannitol | 24.90 |
| D-(+)-Mannose | 16.39 |
| Melibiose | 11.97 |
| D-(+)-Melezitose | 9.94 |
| D-(+)-Raffinose | 10.28 |
| L-(+)-Rhamnose | 14.47 |
| D-(-)-Ribose | 33.48 |
| Salicin | 26.81 |
| D-Sorbitol | 35.97 |
| Stachyose | 9.72 |
| Sucrose | 11.00 |
| Trehalose | 11.01 |
| Xylitol | 32.38 |
| D-(+)-Xylose | 13.84 |

|
Column:
Rezex RPM-Monosaccharide
Bio-Rad® Aminex® HPX-87P |
|
|
Dimension:
300 x 7.8 mm
|
|
|
Mobile Phase:
Water
|
|
|
Flow Rate:
0.6 mL/min
|
|
|
Temperature:
80 °C
|
|
|
Detection:
ELSD
|
|
|
Sample:
1. Glucose
2. Erythritol 3. Mannitol 4. Xylitol 5. Sorbitol |

|
Column:
Rezex RPM-Monosaccharide
Aminex HPX-87P |
|
|
Dimension:
300 x 7.8 mm
|
|
|
Mobile Phase:
Water
|
|
|
Flow Rate:
0.6 mL/min
|
|
|
Temperature:
75 °C
|
|
|
Detection:
ELSD
|
|
|
Sample:
1. Melezitose
2. Maltose 3. Glucose 4. Mannose 5. Fructose 6. Ribitol |

|
Column:
Rezex RPM-Monosaccharide
|
|
|
Dimension:
300 x 7.8 mm
|
|
|
Part No.:
00H-0135-K0
|
|
|
Mobile Phase:
Water
|
|
|
Flow Rate:
0.6 mL/min
|
|
|
Temperature:
85 °C
|
|
|
Detection:
ELSD
|
|
|
Sample:
1. D-Cellobiose
2. α-D-glucose 3. D-(+)-Xylose 4. D-(+)-Galactose 5. D-(-)-Arabinose |

|
Column:
Rezex RPM-Monosaccharide
|
|
|
Dimension:
300 x 7.8 mm
|
|
|
Part No.:
00H-0135-K0
|
|
|
Mobile Phase:
Water
|
|
|
Flow Rate:
0.6 mL/min
|
|
|
Temperature:
75 °C
|
|
|
Detection:
RI
|
|
|
Sample:
1. Stachyose
2. Maltose 3. Glucose 4. Xylose 5. Galactose 6. Fructose 7. meso-Erythritol |
8. Mannitol 9. Salicin 10. Xylitol 11. Sorbitol |

|
Column:
Rezex RPM-Monosaccharide
|
|
|
Dimension:
100 x 7.8 mm
|
|
|
Part No.:
00D-0135-K0
|
|
|
Mobile Phase:
Water
|
|
|
Flow Rate:
0.6 mL/min
|
|
|
Temperature:
80 °C
|
|
|
Detection:
ELSD (Ambient)
|
|
|
Sample:
1. Mannitol
2. Sorbitol |
| Compound | App ID No. |
|---|---|
| D-Cellobiose | 14967 |
| D-(+)-Galactose | 14967 |
| D-(+)-Xylose | 14967 |
| D-(+)-Arabinose | 14967 |
| Erythritol | 14968 |
| Fructose | 5508, 14985 |
| Galactose | 5508 |
| Glucose | 5508, 14985 |
| Maltose | 5508, 14985 |
| Mannitol | 5508, 14968 |
| Compound | App ID No. |
|---|---|
| Mannose | 14985 |
| Melezitose | 14985 |
| meso-Erythritol | 5508 |
| Ribitol | 14985 |
| Salacin | 5508 |
| Sorbitol | 5508, 14968 |
| Stachyose | 5508 |
| Xylitol | 5508, 14968 |
| Xylose | 5508 |
| Rezex RPM-Monosaccharide |   |
|---|---|
| Part No. | 00H-0135-K0 |
| Ionic Form | Lead |
| Standard Dimensions | 300 x 7.8 mm |
| Matrix | Sulfonated Styrene Divinyl Benzene |
| Cross Linking | 8 % |
| Particle Size (μm) | 8 |
| Min. Efficiency (p/m) based on last peak | 35,000 |
| Typical Pressure (psi @ Testing Flow Rate) | 190 |
| Max. Pressure (psi @ Max Flow Rate) | 1,000 |
| Max. Flow Rate (mL/min) | 1.0 |
| Max. Temperature (°C) | 85 |
| Typical Mobile Phase | Water |
| pH Range | Neutral |
| Guard Column Part | 03B-0135-K0 |
| Cleaning, Regeneration and Storage |   |
|---|---|
| Inorganic Modifiers | 5 % Lead Nitrile |
| Organic Modifiers (Max) | 5% Methanol, IPA, EtOH |
| Avoid | Acids, Bases, Non-Lead Salts or Metal Ions, > 30 % Acetonitrile |
| Cleaning Solvent | 100 % Water |
| Flow Rate(mL/min) | 0.4 |
| Temperature (°C) | 85 |
| Duration (hrs) | 12 |
| Regeneration Solvent | 0.1 M Pb(NO3)2 |
| Flow Rate (mL/min) | 0.2 |
| Duration (hrs) | 4-16 |
| Ship/Storage Solvent | Water |