EPA Method 555B, Chlorinated Acids
LC Conditions
Official Method
EPA 555
Column
Synergi 4 µm Hydro-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4375-E0
Phase Name
Hydro-RP
Elution Type
Gradient
Mobile Phase
A: 25mM Phosphoric Acid
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
20
10
90
3
30
10
90
Flow Rate
1.50 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, AQ C18 4 x 3.0mm, 10/Pk Part No.: AJ0-7511 - Holder Part No.: KJ0-4282