ABN Mix on Synergi 4u Hydro-RP 150x4.6

ABN Mix on Synergi 4u Hydro-RP 150x4.6
LC Conditions
Column

Synergi 4 µm Hydro-RP 80 Å, LC Column 150 x 4.6 mm, Ea

Brand

Synergi

Part No

00F-4375-E0

Phase Name

Hydro-RP

Elution Type

Gradient

Mobile Phase

A: Phosphate Buffer, pH 2.5

B: Acetonitrile

C: 0

D: 0

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

8

20

80

3

10

20

80

4

10.5

95

5

5

12

95

5

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Diode Array (PDA) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, AQ C18 4 x 3.0mm, 10/Pk Part No.: AJ0-7511 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Synergi 4 µm Hydro-RP 80 Å, LC Column 150 x 4.6 mm, Ea
00F-4375-E0

Synergi 4 µm Hydro-RP 80 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Pyridine
  • 2. Quinidine
  • 3. Sulfathiazole
  • 4. Triprolidine
  • 5. Benzyl alcohol
  • 6. Nortryptyline
  • 7. Phenol
  • 8. 3-Methyl 4-benzoic acid
  • 9. 3-Methyl salicylaldehyde
Similar Applications