Amphetamines by LC/MS/MS using Strata-X-Drug B and a Kinetex C18 2.6µm, 50x2.1mm
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid
B: Methanol with 0.1% Formic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
1
30
70
3
3
30
70
4
3
30
70
Flow Rate
400.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk
Part No
8B-S128-UCH
Pretreatment Note
1. Before blowing down eluent add 300µL of 0.5N methanolic hydrochloride (to prevent loss of analyte) 2. Dilute sample by a factor of 20 before injecting onto LCMS (Reconstitute in 10% methanol in formic acid (0.1%)).
Load
Load 4.00 mL Sample diluted in 100mM Phosphate buffer
Wash 1
2mL of 100mM sodium acetate buffer pH 5.0
Wash 2
2mL of Methanol
Dry
10 min at 10 inches of Hg
Elution 1
2mL of Ethyl acetate: IPA:NH4OH (7:2:1)
Evaporation
Evaporate to dryness
Reconstitute
1000 µL 10% methanol in formic acid (0.1%)
Order Items Used in This Application
- 1. Amphetamine-d11
- 2. Amphetamine
- 3. Methamphetamine-d14
- 4. Methamphetamine
- 5. MDA-d5
- 6. MDA
- 7. MDMA-d5
- 8. MDMA
- 9. MDEA-d5
- 10. MDEA
Similar Applications
No Similar applications found.