Amphetamines by LC/MS/MS using Strata-X-Drug B and a Kinetex C18 2.6µm, 50x2.1mm

Amphetamines by LC/MS/MS using Strata-X-Drug B and a Kinetex C18 2.6µm, 50x2.1mm
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic acid

B: Methanol with 0.1% Formic acid

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

1

30

70

3

3

30

70

4

3

30

70

Flow Rate

400.00 mL/min

Temperature

0°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info
Sample Preparation Details
Description

Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk

Part No

8B-S128-UCH

Pretreatment Note

1. Before blowing down eluent add 300µL of 0.5N methanolic hydrochloride (to prevent loss of analyte) 2. Dilute sample by a factor of 20 before injecting onto LCMS (Reconstitute in 10% methanol in formic acid (0.1%)).

Load

Load 4.00 mL Sample diluted in 100mM Phosphate buffer

Wash 1

2mL of 100mM sodium acetate buffer pH 5.0

Wash 2

2mL of Methanol

Dry

10 min at 10 inches of Hg

Elution 1

2mL of Ethyl acetate: IPA:NH4OH (7:2:1)

Evaporation

Evaporate to dryness

Reconstitute

1000 µL 10% methanol in formic acid (0.1%)

Order Items Used in This Application

Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk
8B-S128-UCH

Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk

View Product

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. Amphetamine-d11
  • 2. Amphetamine
  • 3. Methamphetamine-d14
  • 4. Methamphetamine
  • 5. MDA-d5
  • 6. MDA
  • 7. MDMA-d5
  • 8. MDMA
  • 9. MDEA-d5
  • 10. MDEA
Similar Applications

No Similar applications found.