PBDEs including BDE-209 on Zebron ZB-SemiVolatiles

PBDEs including BDE-209 on Zebron ZB-SemiVolatiles
GC Conditions
Column

Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea

Part No

7FD-G027-08

Phase Name

ZB-SemiVolatiles

Injection

Splitless @ 85°C

Oven Program

70 °C for 1.25 min to 240 °C @ 20 °C/min to 320 °C @ 50 °C/min for 18 min.

Carrier Gas

Constant Flow Helium, 0.9 mL/min

Detection

High Res Mass Spec (HRMS) @ (325°C)

Detection Info
Sample Notes

Used a PTV in Solvent Vent Mode with temperature program to 320 in 2 min. Application courtesy of Panu Rantakokko @ National Institute for Health and Welfare, Finland

Order Items Used in This Application

Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
7FD-G027-08

Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea

View Product

  • 1. 13C-BDE-28
  • 2. BDE-28
  • 3. BDE-75
  • 4. BDE-71
  • 5. 13C-BDE-47
  • 6. BDE-47
  • 7. BDE-66
  • 8. 13C-BDE-77
  • 9. BDE-77
  • 10. 13C-BDE-100
  • 11. BDE-100
  • 12. BDE-119
  • 13. 13C-BDE-99
  • 14. BDE-99
  • 15. BDE-85
  • 16. 13C-BDE-154
  • 17. BDE-154
  • 18. 13C-BDE-153
  • 19. BDE-153
  • 20. BDE-138
  • 21. 13C-BDE-183
  • 22. BDE-190
  • 23. 13C-BDE-209
  • 24. BDE-209
Similar Applications