Pesticides in Poultry Meat by GC/ECD on ZB-MultiResidue-1

Pesticides in Poultry Meat by GC/ECD on ZB-MultiResidue-1
GC Conditions
Column

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea

Part No

7HM-G016-17

Phase Name

ZB-MultiResidue-1

Injection

Splitless @ 250°C

Oven Program

150 °C for 0.5 min to 20 °C/min to 220 °C hold 0 min, 6 °C, to 310 °C hold time: 0 min

Carrier Gas

Ramp Flow Helium, 3.4 mL/min

Detection

Electron Capture (ECD) @ (350°C)

Detection Info
Sample Notes

Flow ramp program: 3.4ml/min ramped flow- Hold 6.00 min, ramp 5.00 mL/min to 8.00 mL/min, hold 12.08 min, total 19.00 min

Order Items Used in This Application

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea
7HM-G016-17

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea

View Product

  • 1. 2,4,5,6-tetrachloro-m-xylene (Internal Standard 1)
  • 2. aBHC
  • 3. HCB
  • 4. Lindane
  • 5. Heptachlor
  • 6. Drusban
  • 7. Aldrin
  • 8. Heptachlor Epoxide
  • 9. p,p'-DDE
  • 10. Dieldrin
  • 11. o,p-DDT
  • 12. p,p'-TDE
  • 13. Endrin
  • 14. p,p'-DDT
  • 15. Methoxychlor
  • 16. Merex
  • 17. Decachlorobiphenyl (Internal Standard 2)