ASTM 6730 on ZB-PONA 100m x 0.25mm x 0.5um with ZB-PONA tuning 5m x 0.25mm x 1.0mm

ASTM 6730 on ZB-PONA 100m x 0.25mm x 0.5um with ZB-PONA tuning 5m x 0.25mm x 1.0mm
GC Conditions
Column

Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea

Part No

7MG-G042-17

Phase Name

ZB-DHA-PONA

Guard Column

7AG-G042-22

Injection

Split 150:1 @ 200°C

Inlet Liner

AG2-0A13-01

Oven Program

30°C for 8.5 min to 48°C @ 22°C/min 48°C for 27 min to 141°C @ 3°C/min 141°C for 1 min to 275°C @ 1°C/min 275°C for 2 min

Carrier Gas

Constant Flow Hydrogen, 2.0 mL/min

System

Agilent Technologies 6890 HP 6890

Detection

GC/FID @ (275°C)

Detection Info

Order Items Used in This Application

Zebron PLUS Liner for Agilent & Thermo, 4mm ID Single Taper Z-Liner, EA
AG2-0A13-01

Zebron PLUS Liner for Agilent & Thermo, 4mm ID Single Taper Z-Liner, EA

View Product

Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea
7MG-G042-17

Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea

View Product

Zebron ZB-DHA-PONA-TUNE, GC Cap. Column 5m x 0.25mm x 1µm, Ea
7AG-G042-22

Zebron ZB-DHA-PONA-TUNE, GC Cap. Column 5m x 0.25mm x 1µm, Ea

View Product

  • 1. Ethanol
  • 2. C5 (n-pentane)
  • 3. 2-Methylbutane
  • 4. Tert-Butanol
  • 5. 2,3-Dimethylbutane
  • 6. Methyl tert-butyl ether (MTBE)
  • 7. C6 (n-hexane)
  • 8. 1-Methylcyclopentene
  • 9. Benzene
  • 10. Cyclohexane
  • 11. 3-Ethylpentane
  • 12. trans-1,2-Dimethylcyclopentane
  • 13. C7 (n-heptane)
  • 14. 2,3,3-Trimethylpentane
  • 15. Toluene
  • 16. C8 (n-octane)
  • 17. Ethylbenzene
  • 18. 2,3-Dimethylheptane
  • 19. p-Xylene
  • 20. C9 (n-nonane)
  • 21. 5-Methylnonane
  • 22. 1-Methyl-2-ethylbenzene
  • 23. C10 (n-decane)
  • 24. C11 (undecane)
  • 25. 1,2,3,5-Tetramethylbenzene
  • 26. Naphthalene
  • 27. C12 (dodecane)
  • 28. 1-Methylnapthalene
  • 29. C13 (Tridecane)
Similar Applications