DMSO and Ethanol
LC Conditions
Column
Rezex ROA-Organic Acid H+ (8%), LC Column 150 x 7.8 mm, Ea
Brand
Rezex
Part No
00F-0138-K0
Phase Name
ROA-Organic Acid H+ (8%)
Elution Type
Isocratic
Mobile Phase
A: 0.005 M H2SO4
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
Flow Rate
0.60 mL/min
Temperature
40°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Carbo-H 4 x 3.0mm, 10/Pk Part No.: AJ0-4490 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Ethanol
- 2. Dimethyl sulfoxide
-OEChem-06211003142D<br><br> 9 8 0 0 0 0 0 0 0999 V2000<br> 2.5369 -0.2500 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.4030 0.2500 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.2690 -0.2500 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.8015 0.7249 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.0044 0.7249 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.9590 -0.7869 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.8059 -0.5600 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.5790 0.2869 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.0000 0.0600 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 1 2 1 0 0 0 0<br> 1 9 1 0 0 0 0<br> 2 3 1 0 0 0 0<br> 2 4 1 0 0 0 0<br> 2 5 1 0 0 0 0<br> 3 6 1 0 0 0 0<br> 3 7 1 0 0 0 0<br> 3 8 1 0 0 0 0<br>M END
Compound Name
Ethanol
CID
702
Molecular Formula
C2H6O
Molecular Weight
46
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Similar Applications
Improved Performance
Separation of Residual Solvent Critical Pairs in Less than 15 min using Zebron ZB-624PLUS GC column (APP ID 24776)
Zebron ZB-624plus, GC Cap. Column 30m x 0.32mm x 1.8um, Ea
25983
Chiral
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Cellulose-2 in NP
Lux 5 µm Cellulose-2, LC Column 250 x 4.6 mm, Ea
20496
Chiral
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Cellulose-1 in NP
Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
20476
Chiral
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Cellulose-3 in NP
Lux 5 µm Cellulose-3, LC Column 250 x 4.6 mm, Ea
20383
Chiral
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Cellulose-4 in NP
Lux 5 µm Cellulose-4, LC Column 250 x 4.6 mm, Ea
20339
Chiral
1-Phenylethanol on Lux 5µm Cellulose-1 in RP
Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
20134
Chiral
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Cellulose-1 in RP
Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
20132
1-(9-Anthryl)-2,2,2-Trifluoroethanol on Lux 5µm Amylose-2 in NP
20086
Standard
USP Glycerin Purity Assay on ZB-624
Zebron ZB-624, GC Cap. Column 30 m x 0.53 mm x 3.00 µm, Ea
19292
Application Specific
USP Glycerin Purity Assay on ZB-BAC-2
Zebron ZB-BAC-2, GC Cap. Column 30 m x 0.53 mm x 2.00 µm, Ea
19291
Application Specific
USP Glycerin Purity Assay on ZB-BAC-1
Zebron ZB-BAC-1, GC Cap. Column 30 m x 0.53 mm x 3.00 µm, Ea
19290
Improved Performance
USP 467 Procedure B - Class 2 Mix A by GC/FID on ZB-WAXplus
Zebron ZB-WAXplus, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
16353
Standard
USP 467 Procedure A - Class 2 Mix A by GC/FID on ZB-624
Zebron ZB-624, GC Cap. Column 30 m x 0.32 mm x 1.80 µm, Ea
16349
Standard
Common Solvents on ZB-624
Zebron ZB-624, GC Cap. Column 20 m x 0.18 mm x 1.00 µm, Ea
15954
Standard
Common Solvents on ZB-FFAP
Zebron ZB-FFAP, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
15953
Standard
Common Solvents on ZB-WAX
Zebron ZB-WAX, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
15952
Common Solvents on ZB-1MS
15951
Improved Performance
Solvents on ZB-WAXplus 30x0.25x0.25
Zebron ZB-WAXplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
15926
Gel Permeation Chromatography (GPC)
Distearin in Palm Oil on Phenogel 5u 100A
Phenogel 5 µm 100 Å, LC Column 300 x 7.8 mm, Ea
8330
Ion Exclusion
DMSO and Ethanol
Rezex ROA-Organic Acid H+ (8%), LC Column 150 x 7.8 mm, Ea
5513
Ion Exclusion
Alcohols
Rezex RHM-Monosaccharide H+ (8%), LC Column 300 x 7.8 mm, Ea
5512