FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
THC and Metabolites from Whole Blood on Strata-X-Drug B and Kinetex C18 by LC-MS-MS
19947
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Psychotropic Drugs
Forensics/Toxicology
Chemical Substance
Therapeutic / Clinical

Applications

THC and Metabolites from Whole Blood on Strata-X-Drug B and Kinetex C18 by LC-MS-MS
Analytes
1THC-OH
2THC-OH-d3
3THC-COOH
4THC-COOH-d3
5THC
6THC-d3
Compound Name
Hexestrol Dibutyrate
CID
644022
Molecular Formula
C21H30O3
Molecular Weight
330
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 19947)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 1 mM ammonium formate with 0.1% FA
B: 0.1% FA/methanol:acetonitrile(1:1)
Gradient Profile:
Step No.Time(min)%A%B
105050
23595
34.5595
44.515050
565050
Flow Rate:
0.4 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 5000™
Sample Note:
Mass spec conditions: Polarity Analyte Q1 Q3 DP CE (+) THC-OH (*) 331.0 193.3 55 31 (+) THC-OH 331.0 201.2 55 33 (+) THC-OH 331.0 175.3 55 31 (+) THC-OH-d3 334.0 196.3 55 31 (+) THC (*) 315.2 193.1 60 31 (+) THC 315.2 259.1 60 35 (+) THC-d3 318.2 196.1 60 25 (+) THC-COOH (*) 345.2 327.0 55 29 (+) THC-COOH 345.2 299.2 55 49 (+) THC-COOH-d3 348.0 330.1 55 29 CAD: 7 GS1: 60 GS2: 45 TEM: 600 CUR: 25 IS: 5500 * indicates quant ion
Sample Preparation Details
Description:
Strata-X-Drug B 33 µm Polymeric Strong Cation, 60 mg / 6 mL, Tubes, 30/Pk
Part Number:
Pre Treatment Note:
Evaporate eluant to dryness under 50 °C nitrogen. Reconstitute in 500uL of initial mobile phase. Extraction was automated and performed using a Gilson GX-274 ASPEC system
Sample PreTreatment:
Evaporate eluant to dryness under 50 °C nitrogen. Reconstitute in 500uL of initial mobile phase. Extraction was automated and performed using a Gilson GX-274 ASPEC system
Wash 1:
2mL of ACN:Water (15:85)
Evaporation:
Evaporate to dryness
1
Shake
2mL of Ethyl acetate:Isopropanol (85:15)
1
Wash
2mL of ACN:Water (15:85)

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 3 µm Polar C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
Standard
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific