Abstract

Size-exclusion chromatography is commonly used to quantify monoclonal antibody aggregates, but non-specific electrostatic and hydrophobic interactions can compromise peak shape, recovery, and separation. This technical note evaluates the effect of potassium chloride concentration on two monoclonal antibodies using a sub-2 µm Yarra SEC-X300 column with potassium phosphate buffer. The two antibodies responded differently to changes in ionic strength. One required a moderate potassium chloride concentration to achieve acceptable peak symmetry, with the best result obtained at 200 mM, while excessive salt reduced performance through an attributed salting-out effect. The second antibody produced reasonable peak shape without added salt but showed incremental improvements as potassium chloride increased. These findings demonstrate the importance of optimizing buffer composition for individual antibodies. For platform method development across multiple mAbs, the study recommends 50 mM potassium phosphate with 250 mM potassium chloride at pH 6.8 as a starting condition.

Key Benefits

Introduction

Size exclusion chromatography (SEC) is a non-adsorptive separation that fractionates analytes based upon hydrodynamic volume. Because it is non-denaturing, it is the most common method for determination of protein aggregate in a sample. For protein therapeutics, quantitation cal quality attribute (CQA) since aggregate can have lower efficacy and potentially be immunogenic.

Ideally, aggregate analysis by SEC is done in the protein formulation buffer. However, it is critical that an SEC separation be completely “entropic”- that is, reducing or eliminating electrostatic and hydrophobic interactions that might occur between the protein and the functionalized SEC silica column. Buffer formulations for the protein therapeutic, depending on ionic strength and composition, might not have optimal chromatography.

This is especially true with monoclonal antibodies (mAbs), which are the most common protein therapeutic platform, with approximately 70 mAbs predicted to be in the global market by 2020.1 mAb characterization is complicated by the varied physicochemical properties of these complex, 150 kD proteins- isoelectric points vary, typically 7-9. Relative hydrophobic index also is variable and can be unpredictable based upon many factors like buffer formulation and protein concentration.2

For some mAbs, a formulation like 1X Phosphate Buffered Saline might be an acceptable mobile phase. However, recent work by Goyon and colleagues showed the effects of sodium and potassium additives in SEC, and how salt concentration, as well as buffer type and concentration, can affect separation and aggregate recovery significantly.3 With advances in chromatography, sub-2 μm are becoming the common not only in small molecule, but large molecule as well. SEC is no exception, as throughput demands and characterization needs increase. mAb aggregate is often at very low levels (<0.1 % by peak area compared to monomer) and fragment separation is also a method requirement. In these instances, the high efficiency, sub-2 μm SEC column can increase sensitivity of the method and improve separation of closely eluting fragments. This is of course, provided optimal mobile phase conditions are used.

In this technical note, two monoclonal antibodies are analyzed using sub-2 μm SEC columns using optimized mobile phase conditions. Modulation of the salt in the mobile phase showed improvements in peak shape and recovery.

Materials and Methods

Reagents and Chemicals

Monoclonal antibodies samples were graciously gifted from a customer. Potassium Chloride and Potassium Phosphate (monobasic and dibasic) were purchased from Sigma Aldrich (St. Louis, MO, USA). Mobile phase was filtered using a 0.2 μm Phenex™ Nylon Filtration Membran

Experimental Conditions

HPLC Conditions

HPLC was performed using Yarra™ 1.8 μm SEC-X300 run on an Agilent® 1100 equipped with a UV-Vis Detector. Because minimizing system dwell volume is critical for high performance with sub-2 μm SEC columns, it is important to note that the system used was equipped with a 2 μL micro flow cell (G1315-60024) and was plumbed with 0.005” ID PEEK tubing.

Preliminary mobile phase conditions started with the standard United States Pharmacopeia method of 0.2 M KH2PO4 250 mM KCl, pH 6.2. However, peak shape was considerably better using 50 mM KH2PO4 at pH 6.8 (data not shown). From there, KCl amounts were varied as indicated in Table 1.

Figure 1: mAb 1, SEC Profiles

Columns:
Yarra 1.8 μm SEC-X300
Dimensions:
300 x 4.6 mm
Mobile Phase:
50 mM KH2PO4, pH 6.8 KCl (as indicated)
Part Number:
00H-4745-E0-SS
Flow Rate:
0.3 mL/min
Temperature:
Ambient
Detection:
UV-Vis @ 280 nm
Sample:
Agilent 1100

Table 1. mAb 1, Results

dark-gray text-white
Salt Concentration (mM KCl)
1
1
false
center
top
Retention Time (t)
1
1
false
center
top
Peak Width
1
1
false
center
top
Peak Symmetry
1
1
false
center
top
0
1
1
false
center
top
9.46
1
1
false
center
top
0.445
1
1
false
center
top
0.215
1
1
false
center
top
50
1
1
false
center
top
9.47
1
1
false
center
top
0.4374
1
1
false
center
top
0.223
1
1
false
center
top
150
1
1
false
center
top
9.02
1
1
false
center
top
0.1665
1
1
false
center
top
0.764
1
1
false
center
top
200
1
1
false
center
top
9.10
1
1
false
center
top
0.1557
1
1
false
center
top
0.788
1
1
false
center
top
250
1
1
false
center
top
9.15
1
1
false
center
top
0.1594
1
1
false
center
top
0.785
1
1
false
center
top
300
1
1
false
center
top
9.52
1
1
false
center
top
0.3069
1
1
false
center
top
0.193
1
1
false
center
top

Figure 2. mAb 2, SEC Profiles

APP ID 24482

Figure 3. mAb 2, SEC Profile

APP ID 24483

Table 2. mAb 2, Results

dark-gray text-white
Salt Concentration (mM KCl)
1
1
false
center
top
Retention Time (t)
1
1
false
center
top
Peak Width
1
1
false
center
top
Peak Symmetry
1
1
false
center
top
0
1
1
false
center
top
9.63
1
1
false
center
top
0.1472
1
1
false
center
top
0.661
1
1
false
center
top
50
1
1
false
center
top
9.60
1
1
false
center
top
0.1718
1
1
false
center
top
0.78
1
1
false
center
top
150
1
1
false
center
top
9.77
1
1
false
center
top
0.1753
1
1
false
center
top
0.785
1
1
false
center
top
200
1
1
false
center
top
9.82
1
1
false
center
top
0.1767
1
1
false
center
top
0.788
1
1
false
center
top
250
1
1
false
center
top
9.91
1
1
false
center
top
0.1822
1
1
false
center
top
0.778
1
1
false
center
top
300
1
1
false
center
top
9.96
1
1
false
center
top
0.1745
1
1
false
center
top
0.784
1
1
false
center
top

Results and Discussion

The results from mAb 1 clearly show how modulation of salt can affect peak shape. In Figure 1, we can see increases in salt concentration affect peak shape and chromatography. With no potassium chloride, peak shape is unacceptable. It isn’t until 150 mM potassium chloride are peak symmetries within a reasonable value of 0.764. However, with the addition of more potassium chloride past 200 mM, there is a decrease in performance, with 300 mM showing even worse peak shape than no salt. This can be attributed to the addition of more chloride causing a “salting out” effect.

That is, the mAb is now more hydrophobic, leading to secondary interactions with the stationary phase, causing the peak tailing. For mAb 1,200 mM gives the superior peak shape. mAb 2 is an example of a more well-behaved mAb. That is, even with no potassium chloride in the mobile phase, peak shapes look reasonably good. The addition of even 50 mM potassium chloride gives peak symmetry of 0.78, with incremental improvements with each subsequent increase in potassium chloride concentration. In Figure 2, we do a see a small increase in retention time from 50 mM to 300 mM, indicating slight hydrophobic retention, but not to the determinant of overall peak shape.

In this case, it is more prudent to then look visually at chromatography. Figure 3 shows an overlay with a zoomed in view of monomer and separation of pre-peak and high molecular weight aggregate. In this instance, there is an improvement between separation of pre-peak and monomer.

Conclusion

When developing a method for aggregate analysis of mAb’s by SEC, it is critical to optimize mobile phase conditions to prevent non-specific secondary interactions. Here, the effect of salt is investigated for two different biosimilar mAb’s. The first mAb required a moderate amount of salt for acceptable peak shape. The second mAb performed well even with no salt. However, increases in salt showed incremental improvements in peak shape.

Ideally, buffer and salt concentration are optimized, based upon the requirements for the method. However, when there is a need for a platform method- a general method used to evaluate several different mAb’s- a good starting point for method development would be 50 mM KH2PO4 250 mM KCl, pH 6.8.

FAQs

Why should the mobile phase be optimized for mAb aggregate analysis by SEC?

SEC is intended to separate proteins according to hydrodynamic volume, but electrostatic and hydrophobic interactions between a monoclonal antibody and the stationary phase can alter the separation. These secondary interactions may affect peak shape, aggregate recovery, retention, and the resolution of closely eluting fragments. Because mAbs vary in isoelectric point and hydrophobicity, a formulation buffer that performs well for one molecule may not provide suitable chromatography for another.

How did potassium chloride concentration affect the first monoclonal antibody?

The first mAb produced unacceptable peak shape without potassium chloride. Peak symmetry reached a reasonable value at 150 mM potassium chloride and improved further at 200 mM, which provided the best peak shape under the evaluated conditions. Increasing the concentration beyond 200 mM reduced performance. At 300 mM, the peak shape was worse than with no added salt, which the authors attributed to a salting-out effect and increased hydrophobic interaction with the stationary phase.

Do all monoclonal antibodies require the same salt concentration for SEC?

No. The second mAb produced reasonably good peak shape even without potassium chloride. Adding 50 mM potassium chloride produced a peak-symmetry value of 0.78, followed by incremental improvements as the salt concentration increased. A small increase in retention was also observed from 50 to 300 mM, indicating slight hydrophobic retention, but the document states that this did not adversely affect overall peak shape. These results demonstrate that salt requirements depend on the physicochemical behavior of the individual antibody.

What is a suitable starting mobile phase for a platform mAb SEC method?

When developing a platform method intended to evaluate several different monoclonal antibodies, the technical note recommends beginning with 50 mM potassium phosphate and 250 mM potassium chloride at pH 6.8. This is presented as a method-development starting point rather than a universally optimal condition. Buffer composition and salt concentration should still be optimized according to the peak shape, recovery, and separation requirements of each antibody.

References

  1. Ecker DM, Jones SD, Levine HL. The therapeutic monoclonal antibody market. mAbs. 2015;7(1):9-14. doi:10.4161/19420862.2015.989042.
  2. Haverick M, Mengisen S, Shameem M, Ambrogelly A. Separation of mAbs molecular variants by analytical hydrophobic interaction chromatography HPLC: overview and applications.
    mAbs. 2014;6(4):852-858. doi:10.4161/mabs.28693.
  3. Goyon A, Beck A, Veuthey JL, Guillarme D, Fekete S. Comprehensive study on the effects of sodium and potassium additives in size exclusion chromatographic separations of protein biopharmaceuticals. J Pharm Biomed Anal. 2016. doi: 10.1016/j.jpba.2016.09.031

Release Date: 10/2026 Reviewed Date: 10/2026