Technical Tip: Choosing & Changing Syringe Wash Solvents

Syringe

Scientists aren’t perfect. Nobody is. In our Technical Tips*, we can solve your most debilitating issues—from mechanics to methodology—together.*

We often get queries about why syringes become stuck or jammed with particular methods. Some methods are fine with the same solvent and can last for months, but others require that syringes be replaced every couple of days. What most often happens is the syringe plunger becomes stuck and the autosampler bends the plunger during the injection. After hearing many similar requests, this most often occurs with semi-volatile analytes in volatile solvents like pentane or dichloromethane. The stuck plungers may be a classic case of a contaminated syringe—AKA the syringe is not being effectively cleaned after each injection. In the time between injections, the solvent evaporates and leaves behind concentrated analytes, which become sticky and make the plunger hard to move.

One quick solution may be to use the solvent wash option available with the autosampler. The syringe wash is most effective immediately after the injection to remove as much sample as possible before any solvent from the syringe can evaporate. If already using the solvent wash option, a solution may be increasing the number of times the syringe is washed after the injection.

Changing or adding additional wash solvents can help too, but there are some things to take into consideration: