Biozen dSEC-1 columns are engineered to meet the growing demand for aggregate analysis in peptide therapeutics and mid-sized biologics. Packed with 90 Ã… small pore silica and featuring inert diol-based surface chemistry, dSEC-1 delivers enhanced resolution for low and high molecular weight peptide aggregates. Its robust design ensures reproducibility and compatibility with advanced workflows like SEC-MS and SEC-MALS.
Robust Performance for Peptide Aggregate Analysis
Separation Mode Size Exclusion |
Recommended Use Analysis of peptides, protein fragments, and small biologics with reduced analysis times and improved column lifetimes. |
Column:
Biozen dSEC-1, 1.6 μm
|
Dimension:
150 x 4.6 mm
|
Part No:
00F-4801-E0
|
Mobile Phase (isocratic):
1X PBS:Isopropanol (70:40 v/v)
|
Flow Rate:
0.35 mL/min
|
Injection Volume:
20 µg on column
|
Temperature:
40 ËšC
|
LC System:
Waters H-class
|
Detection:
UV@ 280 nm
|
Sequence Name | Bases | Sequence | MW |
---|---|---|---|
19 mer Oligo std | 19 | TTA GGT CCC AAA AGG AGA C | 5845.9 |
16 mer Oligo std | 16 | GTG CAT CTT CTG AGG T | 4903.2 |
Analytes (Diluent):
ChoITEG conjugated siRNA (Water)
|
Column:
Biozen dSEC-1, 3 µm
|
Dimension:
300 x 4.6 mm
|
Part No.:
00H-4801-E0
|
Temperature:
10 ËšC
|
Flow Rate:
0.10 mL/min
|
Mobile Phase (isocratic):
10 mM Potassium phosphate (pH 7.0), 200 mM Potassium chloride, 40%, Acetonitrile
|
Injection Volume:
1 µL
|
Instrument:
Thermo Vanquish Horizon
|
Wavelength (nm):
UV @ 260 nm
|
Sequence Name | Bases | Sequence | MW |
---|---|---|---|
Sense (S) | 16 | N/A - Chl1TEG | 5900 |
Antisense (A) | 21 | N/A | 6900 |
Analytes (Diluent):
ChoITEG conjugated siRNA (Water)
|
Column:
Biozen dSEC-1, 3 µm
|
Dimension:
300 x 4.6 mm
|
Part No.:
00H-4801-E0
|
Temperature:
10 ËšC
|
Flow Rate:
0.08 mL/min
|
Mobile Phase (isocratic):
10 mM Potassium phosphate (pH 7.0), 200 mM Potassium chloride, 35%, Acetonitrile
|
Injection Volume:
1 µL
|
Instrument:
Thermo Vanquish Horizon
|
Wavelength (nm):
UV @ 260 nm
|