How Much Sample Can I Upload Onto My SPE Device?

Sample Preparation Tip

Level: Basic

Is it linked to the concentration of my analyte of interest?

When considering the mass of sorbent required to perform a solid phase extraction it is critical to consider the sample matrix (serum, plasma, urine, saliva, food, river water, soil etc). This information will allow you, with the use of a suitable look up table, to determine the mass of SPE sorbent to use. The tables are produced to reflect the relative levels of interfering endogenous material present in the samples. The SPE device must have enough capacity to retain all analytes of interest together with any endogenous materials capable of retaining on the sorbent with the same type of interactions as the analyte. If there is insufficient sorbent, it is possible some analyte may pass unretained through the SPE device during sample loading, leading to lower recoveries and a loss of consistency of recovery between samples.

Sorbent Mass
Plasma/ Serum/ Blood
Urine
Saliva
Tissue Homogenates
Water (Particulate Free)
Food/ Plant Materia
Soil

Polymeric Phases: X, X-C, X-CW, X-A, X-AW

10 mg
100 µL

250 µL
250 µL
20 mg

20 mL
100 mg
2 g
30 mg
250 µL
1 mL
1 mL
50 mg
60 mL
300 mg
6 g
60 mg
500 µL
2 mL

2 mL
100 mg

120 mL
600 mg
12 g
100 mg
1 mL
4 mL
4 mL
200 mg
200 mL
1 g
20 g
200 mg
2 mL
8 mL
350 mg
400 mL
2 g
40 g
500 mg
5 mL
20 mL
1g
1L
5g
100g
1 g
2 g
2 L
10g
200g
2 g
4 g
4L
20 g
400 g

Silica Phases: Si-1, C18-E, C18-U, C8, NH2, etc.

25 mg
100 µL
250 µL
250 µL
20 mL
25 mL
125 mg
2.5 g
50 mg
250 µL
500 µL
500 µL
50 mg
50 mL
250 mg
5g
100 mg
500 µL
2 mL
2 mL
100 mg
100 mL
500mg
10g
200 mg
1 mL
4 mL
4 mL
200 mg
200 mL
1 g
20g
500 mg
2 mL
8 mL
500 mg
500 mL
2.5 g
50 g
1 g
5 mL
Column3
20 g
1 g
1 mL
5 g
100 g
2 g
2 g
2 mL
10 g
200 g
5 g
5 g
5 mL
20 g
500 g
10 g
10 g
50 g

Generic Loading Capacities

Polymeric: 10 - 15% of Total Bed Mass | Silica: 5% of Total Bed Mass
If we compare plasma and urine, we can load 500 µl of plasma onto a 60 mg polymeric sorbent such as Strata™-X. Using the same 60 mg SPE tube 2 mL of urine could be loaded, indicating that urine contains significantly less endogenous material than plasma.

For sample types not included in the table, the use of generic loading capacity, 5% of sorbent mass for silica materials, or 10-15% for polymeric materials are advised. The difference between silica and polymeric sorbent loading capacity is a function of available surface area, which is higher for polymeric materials and also ligand density – the entire surface of a polymeric material is essentially stationary phase, while on a silica material only the phase bonded to the silica takes part in the extraction mechanism for reversed phase or ion exchange extractions.