Strata Screen-C Solid Phase Extraction (SPE) Products

Strong Cation-Exchange Solid Phase Extraction (SPE)

Incorporates the hydrophobic selectivity of a C8 phase and strong cation exchange for the extraction of basic drugs from biological matrices.

#ffffff00
290
2090
strata-solid-phase-extraction-products
strata-screen-c
smooth-scroll-vertical-tabs

Available Formats

96-Well-Plate
1, 3, and 6mL Tubes

General Starting Method

Strata Screen-C Strong Cation-Exchange

Condition
1mL Methanol
Equilibrate
1 mL Acidified Water
Load
Pretreated sample (acidified)
Wash 1
1 mL 0.1N HCl in Water
Wash 2
1 mL 0.1N HCl in Methanol, dry under vacuum for 2-5 min
Elute
1 mL 5 % NH4OH in Methanol

Based on 100 mg/1 mL sorbent mass. The above is a convenient starting point for SPE method development, further optimization may be required to tailor the method for your specific extraction needs.