Strata-X-A SPE Products
Strong Anion Mixed-Mode Solid Phase Extraction (SPE)
A strong anion-exchange functionalized polymeric sorbent that allows for complete retention of weakly acidic compounds with a pKa greater than 2, making 100% organic wash conditions possible. A 100% organic wash ensures that the maximum amount of interferences are removed from the target compound.
SP
292
2132
strata-x-solid-phase-extraction-products
strata-x-a
smooth-scroll-vertical-tabs
Order
Overview
3 Mechanism of Retention
3
Strong Anion-Exchange
false
π-π Bonding
false
Hydrophobic Interaction
false
Available Formats
4
96-Well-Plate
false
96-Well-Plate
false
1, 3, and 6mL Tubes
false
Giga Tubes (12-150mL Tubes)
false
Starting Method
For Acids with pKa > 2
Column:
1mL Methanol
Equilibrate:
1 mL Water, pH 6-7
Load:
Diluted Sample pH 6-7
Wash 1:
1 mL 25 mM Ammonium Acetate Buffered, pH 6-7
Wash 2:
1 mL Methanol (collect this fraction to analyze Neutral/Bases)
Elute Acid:
2x 500 µL 5 % Formic Acid in Methanol
Based on 30 mg/1 mL sorbent mass. The above is a convenient starting point for SPE method development, further optimization may be required to tailor the method for your specific extraction needs.