| Products | Zebron GC Columns|Zebron ZB-XLB-HT Inferno|

Zebron ZB-XLB-HT Inferno GC Columns

High Temp Stability, Low Bleed

  • Rugged, non-metal si-arylene GC column stable to 400 ˚C
  • Robust column for high temperature bake outs and analysis, such as high molecular weight compounds
  • Provides unique selectivity for conformational analyses
  • Longer lifetime with high temperature, polyimide coated, fused silica tubing
  • Low activity, provides good peak shape for acidic and basic samples
  • Good tool for general screening to identify unknown samples

Recommended Use
Herbicides/Insecticides, PCBs, Pesticides, Unknown Samples

#ffffff00
GC
15
398
zebron-gc-columns
zebron-zb-xlb-ht-inferno
smooth-scroll-vertical-tabs

Order

Overview

Upgrade to Zebron from these similar* phases:

Aroclor 1242: DIN Method 51527

Column:
Zebron ZB-XLB-HT Inferno
Dimension:
30 meter x 0.25 mm x 0.25 µm
Detection:
ECD @ 350 °C
Part No.:
7HG-G024-11
Injection:
Split 2:1 @ 250 °C, 1 µL, pressure pulse @ 40 psi for first 0.25 min
Carrier:
Helium @ 1.5 mL/min (constant flow)
Oven Program:
50 °C for 0.5 min to 210 °C @ 40 °C/min for 3 min to 230 °C @ 30 °C/min for 5 min to 250 °C @ 30 °C/min for 5 min to 320 °C
@ 40 °C/min for 2 min

Sample:
Total concentration of aroclors was 90 ppm in isooctane.

1. TCMX
2. BZ# 31
3. BZ# 28
4. BZ# 52
5. BZ# 101
6. BZ# 77
7. BZ# 123

8. BZ# 118
9. BZ# 153
10. BZ# 114
11. BZ# 105
12. BZ# 138
13. BZ# 126
14. BZ# 167

15. BZ# 156
16. BZ# 180
17. BZ# 157
18. BZ# 170
19. BZ# 169
20. BZ# 189
21. DCB

Melamine and Cyanuric Acid by GC-MS

Columns:
Zebron ZB-XLB-HT Inferno
Dimension:
15 meter x 0.25 mm x 0.25 µm
Detector:
MSD @ 325 °C
Part No.:
7EG-G024-11
Injection:
On-Column @ 103 °C, 1 µL
Carrier Gas:
Helium @ 1.4 mL/min (constant flow)
Oven Program:
100 °C for 0.5 min to 320 °C @ 25 °C/min

Sample:

Analytes are 200 ng / 100 µL in BSTFA / Pyridine (1:1)
1. Cyanuric Acid 13C3 (IS)
2. Cyanuric Acid
3. Melamine 13C3 15N3 (IS)
4. Melamine

Good Results for Difficult Samples

Explosives by GC-MS

Columns:
Zebron ZB-XLB-HT Inferno
Dimension:
15 meter x 0.25 mm x 0.25 µm
Detector:
MSD @ 325 °C
Part No.:
7EG-G024-11
Injection:
On-Column @ 73 °C, 0.5 µL
Carrier Gas:
Helium @ 1.4 mL/min (constant flow)
Oven Program:
70 °C for 1 min to 140 °C @ 25 °C/min for 4 min to 280 °C @ 25 °C/min

Sample:

Analytes are 10 ppm in dichloromethane

1. Nitrobenzene
2. 2-Nitrotoluene
3. 3-Nitrotoluene
4. 4-Nitrotoluene
5. Nitroglycerin
6. 2,6-Dinitrotoluene (2,6-DNT)
7. 1,3-Dinitrobenzene (1,3-DNB)
8. 2,4-Dinitrotoluene

9. 2,4,6-Trinitrotoluene (2,4,6-TNT)
10. PETN
11. 1,3,5-Trinitrobenzene (1,3,5-TNB
12. RDX
13. 4-Amino-2,6-dinitrotoluene
14. 3,5-Nitroaniline
15. 2-Amino-4,6-dinitrotoluene
16. Tetryl

18783, 19322, 18785, 18480, 18789, 18784

Applications

View All Applications
/resources/application-library#f-phxphases=ZB-XLB-HT