For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
GC-MS/MS Separation of 2,3,7,8-Tetrachlorodibenzofuran (TCDF) from its isomers on ZB-Dioxin GC column with 5m Zebron Guardian
26056
Separation Mode: Application Specific
Application Specific
ZB-Dioxin
GC
Food & Beverage
Environmental
Chromatography
Chromatographic Property

Applications

GC-MS/MS Separation of 2,3,7,8-Tetrachlorodibenzofuran (TCDF) from its isomers on ZB-Dioxin GC column with 5m Zebron Guardian
Analytes
11,3,6,8-TCDF
22,3,7,8-TCDF
31,2,8,9-TCDF
Compound Name
1,3,6,8-Tcdf
CID
51462
Molecular Formula
C12H4Cl4O
Molecular Weight
305
No. Hydrogen Bond Acceptors
1
Details
GC Conditions (App ID: 26056)
Column:
Part No:
Phase Name:
ZB-Dioxin
Injection:
Splitless @ 330°C, 1.0 μL
Inlet Liner:
Oven Program:
to 220°C @ 30°C/min 220°C for 15.5 min to 240°C @ 2°C/min 240°C for 4.5 min to 270°C @ 5°C/min 270°C for 5 min to 330°C @ 15°C/min 330°C for 7 min
Carrier Gas:
Constant Flow Helium, 1.056 mL/min
System:
Agilent Technologies 7890 Agilent Technologies 7890
Detection:
GC/MS/MS (340°C)

Order Items Used in This Application

Similar Applications
search
Zebron ZB-Dioxin, GC Cap. Column 60m, 0.25mm, 0.20um, Ea
Application Specific
Zebron ZB-Dioxin, GC Cap. Column 60m, 0.25mm, 0.20um, Ea
Application Specific
Zebron ZB-Dioxin, GC Cap. Column 60m, 0.25mm, 0.20um, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 60 m x 0.25 mm x 0.25 µm, Ea
Standard