FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Interferon Alpha intact 5% impurity on Jupiter 3u C18 and Jupiter 5u C4
18073
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Clinical Research
Therapeutic / Clinical

Interferon Alpha intact 5% impurity on Jupiter 3u C18 and Jupiter 5u C4

Analytes
1Intact & 5% impurity
Details
LC Conditions (App ID: 18073)
Column:
Brand Name:
Jupiter
Part No:
Phase Name:
C18
Sample Note:
Application Focus: To demonstrate Jupiter 300 utility for separating folded from unfolded biogeneric proteins Unlike other separation techniques, reversed phase can often visualize differences between intact and unfolded/ mis-folded protein states. Especially with E.Coli produced recombinant proteins, refolding analysis is often required as part of both manufacturing process analysis technology as well as quality control testing of a finished product. In this example Interferon alpha was exposed to chemical unfolding conditions at higher temperature in the presence of DTT (3 mM dithiothreitol (DTT) at 55°C for 30 min) and the chromatographic behavior of the unfolded protein was compared to the intact protein by analysis on the Jupiter 300 3u C18 and Jupiter 300 5u C4 columns. In App ID# 18073 5% of unfolded interferon was added to the intact protein. As one can see from the overlaid chromatograms, both columns could easily detect unfolded impurities lower than 5% with good, rapid resolution from the intact protein in fewer than 15 minutes. This application demonstrates the utility of Jupiter 300 media in doing refold analysis.
Similar Applications
search
No Data