For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Aflatoxins on Kinetex 2.6µm C18 50 x 2.1mm ID with Acquity UPLC MS/MS
18721
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Poisons
Food & Beverage
Food and Beverage (incl. additives
contaminants)
Product / Trade Good

Applications

Aflatoxins on Kinetex 2.6µm C18 50 x 2.1mm ID with Acquity UPLC MS/MS
Analytes
1Aflatoxin M1
2Aflatoxin G2
3Aflatoxin G1
4Aflatoxin B2
5Aflatoxin B1
Compound Name
Aflatoxin M1
CID
23236
Molecular Formula
C17H12O7
Molecular Weight
328
No. Chiral Centers:
2
No. Hydrogen Bond Acceptors
7
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 18721)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid and 1mM NH4 formate in H2O
B: 60/40 MeOH/ACN with 0.1% formic and 1mM NH4 formate
Gradient Profile:
Step No.Time(min)%A%B
106337
214060
31.01595
41.5595
51.516337
62.56337
Flow Rate:
600 µL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (110°C)
Detector Info:
ESI+, MRM mode, Aflatoxin M1- 329.3-273.2 (internal standard) Aflatoxin G2 - 331.3-313.2, Aflatoxin G1 - 329.2-234.3, Aflatoxin B2 - 315.3-287.3, Aflatoxin B1 - 313.2-285.3

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna 3 µm PFP(2) 100 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Prodigy 5 µm ODS-3 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase