For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Simvastatin by HPLC- Luna 5um C18(2) 250x4.6mm
19518
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Antilipemic Agents

Simvastatin by HPLC- Luna 5um C18(2) 250x4.6mm

Analytes
1Lovastatin
2Simvastatin
Details
LC Conditions (App ID: 19518)
Column:
Brand Name:
Luna
Part No:
Phase Name:
C18(2)
Sample Note:
For USP's system suitability requirements, the capacity factor for simvastatin must not be less than 3.0; column efficiency not less than 4500 theoretical plates; tailing factor not more than 2.0; resolution, while not specified according to USP standards, should be adequate at no less than 2.0. On the LUNA 5um C18(2) 250x4.6mm, the capacity factor for simvastatin is 4.6; column efficiency is 15,943; tailing factor is 0.985; and resolution between lovastatin and simvastatin is 9.54.
Similar Applications
search
No Data