1Lovastatin |
---|
2Simvastatin |
1Lovastatin |
---|
2Simvastatin |
Column: |
Brand Name: Luna |
Part No: |
Phase Name: C18(2) |
Sample Note: For USP's system suitability requirements, the capacity factor for simvastatin must not be less than 3.0; column efficiency not less than 4500 theoretical plates; tailing factor not more than 2.0; resolution, while not specified according to USP standards, should be adequate at no less than 2.0. On the LUNA 5um C18(2) 250x4.6mm, the capacity factor for simvastatin is 4.6; column efficiency is 15,943; tailing factor is 0.985; and resolution between lovastatin and simvastatin is 9.54. |