FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Identification and Quantification of Designer Drugs in Urine by LC-MS/MS
25865
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Anesthetics,Analgesics,Excitatory Amino Acid Antagonists
Forensics/Toxicology
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Identification and Quantification of Designer Drugs in Urine by LC-MS/MS

Analytes
1Dehydronorketamine
2Norketamine
3Ketamine
4Methoxetamine
53-MeO-PCE
63-MeO-PCP
Details
LC Conditions (App ID: 25865)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
System:
Sciex Applied Biosystems Q TRAP,  Mass Analyzer:  QQQ-LT, Interface(s): ES/APCI, PI
Sample Note:
To 20 µL urine, add 50 µL internal standard (D5-PCP in acetonitrile). Vortex then centrifuge for 10 min at 15,000 rpm. Dilute supernatant with 930 µL Mobile Phase A. Mix and transfer to an autosampler vial for injection.
Similar Applications
No Data