HPLC Column Lifetime - Knowing When It Is Time to Say Goodbye

HPLC column lifetime

Letting go of your favorite column can be hard to do, so when is it really time to say goodbye? Below are some of the signs to look for that your relationship and your HPLC column lifetime is coming to an end.

High Backpressure:

Increases in backpressure are typically the result of a blockage at the inlet frit of the HPLC column. To remove the blockage in most situations you can reverse the flow of the solvent through the column at a reduced flow rate and flow 100% strong solvent through the column to remove this. If your pressure does not return to normal either the frit is blocked irreversibly or there is a void in your column. Both of situations require the column to be replaced.

HPLC column lifetime - high backpressure

Split Peaks:

Many split peaks are also the result of a blocked frit at the inlet of the column which results in a channeled injection. To resolve this follow the steps outlined above first and foremost. If the split peak remains the likely cause is either incorrect running conditions of your method (working too close to the pKa of your compound or using a strong injection solvent) or your column may be fouled with strongly retained contaminants. If the inlet is irreversibly plugged or the stationary phase fouled with contaminants you will need to replace column.

Column lifetime split peaks

Loss of Resolution:

HPLC columns are consumable items and as such over time you will begin to lose resolution of your critical pairs. Some column regeneration may be possible and each column will have its own individual protocol for this, however once loss of resolution of a critical pair of compounds has been reached and cannot be regenerated, your column has reached the end of it’s natural life for the particular analysis you are doing.

Broad Peak Shape:

Some peak broadening can occur for reasons other than the natural end of the life of a column, these include: injection overload, inadequate buffering, inappropriate method conditions, and strong injection solvents. However, if these elements have been eliminated, peak broadening is one sign your column is ageing. Once your peak shape has deteriorated to the point you can no longer pass system suitability or resolve critical pairs it is time to part ways.

HPLC column lifetime - broad peak shape

Retention Time Shifts:

More often than not retention time shifts are the result of poor method control, however in the case of phase contamination you will see retention time drifts of peaks if your phase is becoming fouled. Not all peaks will be affected equally and if you have eliminated all method concerns from your troubleshooting the likely cause is stationary phase contamination and your column will need to go.

To be certain the column is the problem there are a few things you can test

All of these troubleshooting points are discussed further in our HPLC troubleshooting guide which can be downloaded by clicking here.

HPLC troubleshooting guide