FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Blog Hero Background

Fast and Economical PFAS Extraction Using Strata PFAS - SCIENCE UNFILTERED

Blog
Fast and Economical PFAS Extraction Using Strata PFAS
Download PDF PFAS are a class of highly stable synthetic organic compounds used in a wide variety of industrial and commercial applications. They...

Posted on Sep 6, 2022

PFAS are a class of highly stable synthetic organic compounds used in a wide variety of industrial and commercial applications. They are also highly stable in the environment and strongly bioaccumulate. As a result, they have become ubiquitous throughout the globe and are often referred to in today's media as "Forever Chemicals". Consequently, PFAS levels need to be tested in drinking water and more recently methods have been developed to measure PFAS in other environmental matrices that require more complex clean-up solutions, such as wastewater, soils, and sediments.

Accurate, Precise, and Economical PFAS Extraction Solution

The United States Department of Defense (DOD) is dealing with extensive PFAS contamination owing to the widespread use of PFAS-based Aqueous Film Forming Foam (AFFF) used as fire suppression foams at many military installations. As a result, DOD developed its own PFAS analytical guideline to help solve their installations' unique environmental monitoring and clean-up challenges. The US EPA and DOD have been working jointly to validate EPA 1633 Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous, Solid, Biosolids, and Tissue Samples by LC-MS/MS. A single-lab validated draft method is the result of this collaboration method and calls for the use of a polymeric weak anion exchange (WAX) SPE sorbent in combination with graphitized carbon black (GCB) powder. For water samples, the process involves an initial SPE using a WAX cartridge followed by dGCB or the use of a second cartridge. For solid samples, dGCB is added to an initial ammonium hydroxide wash, followed by the WAX SPE cartridge.

Both methods add time, cost, and chances for increased errors to the clean-up procedure and present the opportunity for loss of analytes and introduction of imprecision.  

Strata PFAS SPE, wherein the two sorbents are contained within a single tube, offers the opportunity for decreased sample processing time and increased accuracy and precision. Strata PFAS is a stacked single cartridge solution with polymeric WAX and GCB sorbents that functions as a traditional Solid Phase Extraction (SPE) cartridge with a built-in polishing step. When comparing different recoveries for a small subset of analytes for a WAX SPE and dSPE GCB method versus Strata PFAS, the recovery is greatly improved using Strata PFAS. The results are shown in Table 1.

A routine Laboratory Control Sample (LCS) was conducted by a commercial testing laboratory highly experienced with PFAS analysis. The LCS had been spiked with all 32 target analytes at 25 μg/L and was analyzed with a batch of field samples to demonstrate method performance and data acceptability. As shown in Table 2, all 32 analytes were well within method recovery limits with an average recovery of 98.8% and a mean recovery of 99.0%, thereby demonstrating acceptability of the use of Strata PFAS in the performance DOD QSM5.1/5.3 as well as draft EPA 1633.

SPE Conditions for Analyte Extraction Using Strata PFAS SPE

Cartridge: Strata PFAS 200 mg WAX/50 mg GCB/ 6 mL)
Part No.:
CS0-9207
Condition 1:
4 mL 0.3 % Ammonium hydroxide
Condition 2:
4 mL Methanol
Equilibrate:
5 mL Water
Load:
Add sample at 4 mL/min
Wash:
2x 4 mL Water
Elute:
2x 4 mL 0.3 % Ammonium hydroxide in
Evaporate:
Methanol to dryness and reconstitute to 1 mL with Methanol/Water (96:4)

Table 1: Recovery Comparisons of WAX SPE and dSPE using GCB vs Strata® PFAS Single Cartridge Method

AnalyteWAX SPE + dSPE GCB
% Recovery
Strata PFAS Stacked Cartridge
% Recovery
13C2-PFDoDA7784.5
13C2-PFTeDA6284.0
PFODA3878.3
PFHxDA6389.3

Table 2: Recovery of QSM 5.3 Target Analytes from a Laboratory Control Sample Using Strata PFAS SPE (WAX/GCB)

AnalyteActual
Concentration
Sample
Result
% RecoveryMethod Recommendation
Limits
Pass/Fail
PFBA25.60022.6408884-135Pass
PFPeA25.60022.1578775-138Pass
PFBS22.64022.3009981-133Pass
4:2-FTS23.92022.0789264-134Pass
PFHxA25.60024.6449680-137Pass
PFPeS24.00021.6999082-132Pass
HFPODA25.60026.33610370-130Pass
PFHpA25.60027.01810680-140Pass
PFHxS24.20024.71310271-131Pass
DONA24.12026.08310870-130Pass
6:2-FTS24.28024.21710051-155Pass
PFHpS24.36023.0159480-129Pass
PFOA25.60025.0439883-138Pass
PFOS24.48022.4929254-139Pass
PFNA25.60025.87210173-140Pass
9Cl-PF3ONS23.84021.8639270-130Pass
PFNS24.56021.9939071-121Pass
PFDA25.60025.0479878-137Pass
8:2-FTS24.52022.2319162-133Pass
PFOSA25.60025.71410073-121Pass
NMEFOSAA25.60030.90612153-136Pass
PFDS24.64022.8739369-124Pass
PFUnDA25.60026.35310370-134Pass
NEtFOSAA25.60028.76511259-145Pass
11Cl-PF3OUdS24.12022.6259470-130Pass
PFDoDA25.60027.71010875-139Pass
10:2-FTS24.68026.62610850-124Pass
PFDoS24.80021.5098739-121Pass
PFTrDA25.60025.81410167-144Pass
PFTeDA25.60025.4469979-134Pass
PFHxDA25.60029.66211636-136Pass
PFODA25.60027.37310710-124Pass
Andrew Patterson1 , Charles Neslund2 , Robert Brown2 , Sam Lodge3 , David Kennedy3 , and Brian Marshall3
1 Eurofins Environment Testing America, 800 Riverside Pkwy, Sacramento, CA, 95605, USA 2 Eurofins Environment Testing America, 2425 New Holland Pike, Lancaster, PA 17601, USA 3 Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501, USA