Installing UHPLC Columns on Waters® Systems
Phenomenex LC columns universally fit into all HPLC/UHPLC systems. It's simple to install Phenomenex UHPLC columns onto Waters® ACQUITY® UPLC® Systems.
Posted on Aug 9, 2017
Did you know that Phenomenex LC columns are a universal fit for all HPLC and UHPLC systems on the market?
Even onto Waters® ACQUITY® UPLC® Systems!
It's simple to install Phenomenex UHPLC columns onto Waters® ACQUITY® UPLC® Systems. Here's how!- Disconnect the pre-column heater from the oven.
- Separated the front and back nuts so that the gold front nut can slide easily to adjust the pilot depth.
- NOTE: If the nut does NOT slide easily, chances are your plastic ferrule is damaged and needs to be replaced. Plastic ferrules should be replaced every 5 to 10 times you connect a column.
- Insert the tubing into the column inlet.
- Tighten finger tight, maintaining firm contact between tubing and column inlet.
- Tighten the back retaining nut.
- Install your outlet tubing using the plastic finger tight ferrule.