Strata-X PRO is a new innovative solid phase extraction (SPE) sorbent that offers a faster, cleaner way to extract your samples, completely revolutionizing traditional SPE methods. Strata-X PRO offers improved and rugged polymeric sorbent performance combined with matrix removal technology for a revolutionary solution. With a faster SPE method, it results in at least 40% reduction in time on your SPE protocol. Less steps with no conditioning or equilibration creates a fast SPE method without losing out on the importance of SPE cleaning up your samples.



Microelution 96-Well-Plate

96-Well-Plate

1, 3, and 6mL Tubes

| SPE Mechanism | Analyte Functional Group | Sorbent Functional Group | |
|---|---|---|---|
| 
             Reversed Phase 
             | 
            
             R  
            hydrocarbon aromatic 
             | 
            
             R  
            hydrocarbon 
            aromatic 
             | 
             ![]()  | 
        

Load:  
        500 µL Pre-treated sample/buffer* (1:1) Apply 2-5” Hg vacuum until liquid is no longer visible above top frit  | 
    
Wash: 
        600 µL 5  % Methanol in Water   | 
    
Elute: 
        600 μL 0.1 % Formic acid in Acetonitrile/Methanol (90:10) Apply 2-5” Hg vacuum for 1 minute  | 
    

Load:  
        1 mL Pre-treated sample/0.1 % Formic acid in Acetonitrile (1:4) Apply 5” Hg vacuum until all tubes or wells have cleared  | 
    
Elute: 
        75 μL Water/0.1 % Formic acid in Acetonitrile (1:4)  Apply 5” Hg vacuum until all tubes or wells have cleared  |