For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
USP Alosetron Tablet Assay Standard/Sample solution (0.0115 mg/mL) on Gemini 3 µm C18 75 x 4.6 mm using Agilent 1260 Binary UHPLC
26385
Separation Mode: Reversed Phase
Reversed Phase
USP
C18
HPLC
Serotonin Antagonists
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

USP Alosetron Tablet Assay Standard/Sample solution (0.0115 mg/mL) on Gemini 3 µm C18 75 x 4.6 mm using Agilent 1260 Binary UHPLC

Analytes
1Alosetron HCl
Compound Name
Lotronex
CID
60758
Molecular Formula
C17H19ClN4O
Molecular Weight
330
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
2
Details
Official Method: USP
LC Conditions (App ID: 26385)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Isocratic
Mobile Phase:
A: 0.02 M monobasic sodium phosphate. Adjust with phosphoric acid to a pH of 3.0
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
110 80 20
Flow Rate:
1 mL/min
Temperature:
25°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/UV (DAD, PDA) @ 220.0000000000 nm (nanometers)

Order Items Used in This Application

Similar Applications