FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Documents
Guard Column Lifetime and Capacity to Protect the Biozen dSEC-2 Column
In this technical note, we evaluate the robustness of the guard and attached analytical column towards the efficiency of the Trastuzumab biosimilar Kanjinti ® at suitable loading mass upon repeatedly injecting high loading masses of bovine serum albumin (BSA). The robustness of the guard alone is also reaffirmed upon performing repeated injections of Uridine dissolved in water.

Guard Column Lifetime and Capacity to Protect the Biozen dSEC-2 Column

In this technical note, we evaluate the robustness of the guard and attached analytical column towards the efficiency of the Trastuzumab biosimilar Kanjinti ® at suitable loading mass upon repeatedly injecting high loading masses of bovine serum albumin (BSA). The robustness of the guard alone is also reaffirmed upon performing repeated injections of Uridine dissolved in water.
Document Type:
Technical Notes
Target Industries:
Techniques:
Brands:
Separation Modes:
Size Exclusion