FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Documents
Meeting System Suitability for USP Amlodipine and Olmesartan Medoxomil Tablets Assay and Organic Impurities
In this application note, we report the separation of Amlodipine and Olmesartan Medoxomil and their related organic impurities per the USP monograph using a Kinetex 5 µm Biphenyl column and a Kinetex 5 µm Phenyl-Hexyl column as alternative L11 columns to the ZORBAX 5 µm SB-Phenyl column originally used to elucidate the monograph

Meeting System Suitability for USP Amlodipine and Olmesartan Medoxomil Tablets Assay and Organic Impurities

In this application note, we report the separation of Amlodipine and Olmesartan Medoxomil and their related organic impurities per the USP monograph using a Kinetex 5 µm Biphenyl column and a Kinetex 5 µm Phenyl-Hexyl column as alternative L11 columns to the ZORBAX 5 µm SB-Phenyl column originally used to elucidate the monograph