The cleaning procedure for Luna NH2 columns will depend on the separation mode it is being for. Before starting any kind of cleaning procedure, make sure your in-column solvent or mobile phase is miscible with the recommended cleaning solvents. Also, flow-rates should be 1/5-1/2 of the typical flow-rate. To estimate column volume, use the following equation: V=πr²L V= column volume in mL r= column radius in cm L= column length in cm See the cleaning protocols below. NH2 for Normal Phase: Rinse with 10 Column Volumes each of: Chloroform Isopropanol Methylene Chloride Mobile Phase NH2 for Reversed Phase: Rinse with 10 Column Volumes each of: 95 % Water/5% Acetonitrile (for buffer removal) THF 95% Acetonitrile/5% Water Mobile Phase NH2 for HILIC or Ion-Exchange*: Rinse with 10 Column Volumes each of: 50/50 Organic (e.g. Acetonitrile or Methanol)/20mM ammonium bicarbonate pH 10 (to clean ionically bound compounds) Water Mobile Phase *This column cleaning procedure should only be done infrequently, as repeated exposure to high pH solutions can cause silica dissolution, resulting in peak shape issues

While both form helical structures that are maintained in coating, the amylose can be described as more tightly coiled. The looser cellulose helical structure may more readily accommodate enantiomer interaction. Practically, this results in different selectivity even if the same chiral selector is substituted, as the 3-dimensional orientation around the CSP will differ.

Traditional logic suggests that the columns should be placed in order of increasing pore size (smallest pore size column closest to the injector). In practice, the order is typically not important as long as it is documented (perhaps experimentally determined) to remain consistent in subsequent analyses.

Below are the part numbers for the appropriate SFC SecurityGuard Holders. These guard cartrudge holders have a Teflon O-rings and a hexnut for wrench tightening. AJ0-8617 SecurityGuard SFC PREP Cartridge Holder Kit for 21.2mm ID SFC Columns AJ0-8618 SecurityGuard SFC PREP Cartridge Holder Kit 30mm ID SFC Columns AQ0-8500 PREP Replacement O-Rings, Teflon® For 15 x 21.2mm SG Holder, Size 2-021 2/Pk AQ0-8501 PREP Replacement O-Rings, Teflon®, For 15 x 30.0mm SG Holder, Size 2-025 2/Pk Please contact us for additional SFC product questions.

21.2 mm and 30 mm ID AXIA columns are compatible with 1/16 tubing and 10/32 fittings. 50 mm ID AXIA columns are compatible with 1/8 tubing, although they are shipped with reducing unions allowing for the use of 1/16 tubing and fittings.

Phenomenex has a variety of preparative C18 columns to choose from. If the you are looking for a material that can handle large sample volumes then Luna C18(2) would be the best option. Gemini NX-C18 offers extended pH stability for high pH analyses and a different selectivity to Luna C18(2). Kinetex core-shell columns are best suited for those who are looking for high efficiency preparative applications often where closely eluting impurities pose a problem. Kinetex C18 is also fully scalable from UHPLC to Prep so it is perfect for labs that require the same selectivity across multiple particle sizes. Luna Omega Polar C18 and PS C18 offer a different selectivity to more traditional C18 phases and would be ideal for analysts who need a different selectivity or greater retention of polar compounds.

If you have a working separation at analytical scale, it is possible to use a simple equation to scale up flow rate, injection volume and mass of sample to allow for a comparable separation at preparative scale. This equation can be used as a guide. Generally it is possible to load more, and loading will normally be increased until the point at which resolution is compromised. F2 = F1 x (L2/L1) X (r2/r1)2 F2 is the new flow rate, F1 the original at analytical scale, L2 the length of the new column, L1 the length of the original, r2 the radius of the new column, r1 the radius of the original. The factor generated by this equation can be applied to flow rate, injection volume and injection mass.

When injecting samples in a solvent with a higher elutropic strength than the mobile phase, there is always the potential for peak fronting to occur. When working on a preparative scale, this may be unavoidable due to sample solubility issues or simply the fact that the sample solvent cannot be switched. The use of prep frit in front of your Axia column can be beneficial in these cases. It can catch impurities that are not soluble in mobile phase, preventing the column from blocking. Additionally it promotes mixing between the strong sample solvent and the mobile phase, this can help to reduce the fronting effect seen due to the introduction of the sample in the stronger solvent.